68 Maintenance and Service Welch Allyn Spot Vital Signs LXi
7. P r e s s  t h e  Navigation button to highlight “Blood Pressure” on the display and press 
the Select button.
8. Press the Navigation button to highlight “BP Calibration Check” on the display and 
press the Select button.
9. Press the Select button to close the valve.
10. Verify that the pressure meter is on and the thumb screw valve is closed. Inflate the 
device manually to about 250 mmHg.
11. Drop the pressure to 200 mmHg, wait 15 seconds for stabilization, and take a 
reading.
12. Repeat for 150 mmHg, 50 mmHg, and 0 mmHg (all measuring downscale).
13. If the calibration at any point is outside of ±3 mmHg, call Welch Allyn Technical 
Service for assistance.
Temperature Calibration Check
Use the 9600 Plus Calibration Tester (see “Temperature” on page 72) to check the 
SureTemp Plus or Braun ThermoScan PRO 4000 thermometer accuracy. If the 
thermometer is out of calibration, contact Technical Service.
Masimo SpO
2
 Calibration Check
Use a Masimo-approved SpO
2
 simulator (Fluke Biotek or Clinical Dynamics SmartSat) to 
check the SpO
2
 accuracy. There is no way to change the calibration of the SpO
2
 module. If 
the SpO
2
 is out of calibration, contact Technical Service. 
Nellcor SpO
2
 Functional Check
Use a Nellcor SpO
2
 simulator (SRC-MAX ) to check the SpO
2
 funcationality. There is no 
way to change the functionality of the SpO
2
 module. If the SpO
2
 is not functioning 
properly, contact Technical Service.
Product Disposal
Do not dispose of this product as unsorted municipal waste. Prepare this product for 
reuse or separate collection as specified by Directive 2002/96/EC of the European 
Parliament and the Council of the European Union on Waste Electronic and Electrical 
Equipment (WEEE). If this product is contaminated, this directive does not apply.
For more specific disposal information, see www.welchallyn.com/weee,or contact Welch 
Allyn Customer Service at +44 207 365 6780.