Product Manual
Directions for Use Maintenance and Service 67
Calibration
Blood Pressure Calibration Check
The calibration check is a simple, yet valuable test to determine that the unit is sensing 
pressure accurately. Verify the pressure measurement accuracy of the Welch Allyn Spot 
LXi with an accurate, calibrated pressure meter or sphygmomanometer.
To perform the calibration check:
Have the following equipment available:
1. Disconnect the blood pressure cuff from the blood pressure tubing.
2. Attach two pieces of the 4” tubing to the T-connector. Verify that the tubing is 
positioned perpendicular to each other.
3. Attach the pressure meter or sphygmomanometer to one of the tubes and the bulb 
and value assembly to the second tube.
4. Push a female fitting into the third piece of 4” tubing and connect the opposite end of 
the tubing to the T-connector.
5. Twist the blood pressure tubing fitting that connects to the right blood pressure cable 
connection port to the female fitting and connect the opposite end of the blood 
pressure tubing to the blood pressure cable connection port. Verify that all 
connections are tight.
Figure 11. Blood Pressure Calibration Tubing Connections
6. Enter the Internal Configuration Mode (see “Internal Configuration” on page 31). 
• Accurate, calibrated pressure meter or 
sphygmomanometer
• 4” tubing with an inside diameter of 
approximately .250” (quantity 3 each)
• T-connector • Bulb and valve (5088-01)
• Female fittings (12P524-1) (quantity 2 each)
Right female 
fitting
Accurate, calibrated 
pressure meter or 
sphygmomanometer
T-connector
Bulb and valve 
4” tubing 
Right blood 
pressure cable 
connection