Maintenance and Service
Cleaning
Spot Vital Signs LXi
As necessary, clean the device with a cloth slightly dampened with either 70% isopropyl 
alcohol, 10% chlorine bleach solution, or mild detergent in water, or use PDI Sani-System 
Cloths. Never immerse Spot LXi in any type of fluid.
Blood Pressure Cuff
Refer to the Directions for Use provided with the blood pressure cuff for cleaning 
procedures.
Blood Pressure Hose and Cable
Wipe the pressure hose with a damp cloth moistened in a mild detergent solution. Do not 
immerse hose.
SureTemp Plus Thermometer
Temperature Probe
Press down on the connector tab and slide the connector out of the port to remove the 
temperature probe.
WARNING Before performing any maintenance or service to the Spot LXi, 
disconnect the AC power line from the electrical outlet.
Caution Prevent water or other fluids from entering any connectors. If the 
connectors get wet, dry them with warm air. Check all measurement functions.
Caution Do not sterilize or autoclave the Spot LXi device.
Caution DO NOT immerse or soak the probe in any type of fluid.
Caution DO NOT use steam, heat, or gas sterilization on the probe. 
Caution DO NOT autoclave the probe.