Product Manual

Directions for Use Specifications 57
SpO
2
Specifications
Masimo Sensor Accuracy Guide
Accuracy specified when used with Masimo SET pulse oximetry monitors or with
licensed Masimo SET pulse oximetry modules using PC series patient cables, during no
motion. Numbers present ± 1 standard deviation. Plus or minus one standard deviation
represents 68% of the population. SpO
2
accuracy from 70% to 100%. Pulse rate
accuracy from 25 to 240 bpm.
Masimo Patents
The Masimo sensors and cables are covered under one or more of the following U.S.A.
patents: 5,758,644; 5,823,950; 6,011,986; 6,157,850; 6,263,222; 6,501,975; and other
applicable patents listed at www.masimo.com/patents.htm.
Performance Measurement Range SpO
2
: 1 to 100%
Pulse Rate: 25 - 240 beats per minute (BPM)
Perfusion 0.02% to 20%
Accuracy Saturation: 70% to 100%
No Motion: Adults, Pediatrics ± 2 digits
Neonates ± 3 digits
Motion: Adults, Pediatrics ± 3 digits
Neonates ± 3 digits
Low Perfusion: Adults, Pediatrics ± 2 digits
Neonates ± 3 digits
Pulse Rate Accuracy Pulse Rate: 25 to 250 bpm
No Motion: Adults, Pediatrics, Neonates ± 3 digits
Motion: Adults, Pediatrics, Neonates ± 5 digits
Low Perfusion: Adults, Pediatrics, Neonates ± 5 digits
Table 25. Masimo Sensor Accuracy Guide
Saturation Accuracy Pulse Rate Accuracy
Sensor Weight Range No Motion Motion No Motion Motion
LNOP-ADT > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm
LNOP-PDT 10 to 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm
LNOP-NEO < 10 kg ± 3% ± 3% ± 3 bpm ± 5 bpm
LNOP-NEOPT < 1 kg ± 3% ± 3% ± 3 bpm ± 5 bpm
LNOP-DCI > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm
LNOP-DCIP 10 to 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm