Product Manual

Directions for Use Controls, Display Window, and Connections 29
Battery
Charge the Spot LXi battery and printer battery (if using) for 6 hours before initial use.
Charge the device an additional hour if it includes a Braun ThermoScan PRO 4000
thermometer.
While Spot LXi is charging, the charging LED (~) flashes and the battery level indicator
segments on the display continuously sequence. When the battery is fully charged, the
charging LED is solid and the battery level indicator is steady with all segments
continuously shown on the display.
If the device includes a Braun ThermoScan PRO 4000 thermometer, the charging LED
below the thermometer will illuminate orange as it is charging. When the battery is fully
charged, the LED will power off.
If the Spot LXi has a mounted, external printer that is connected via the cable, the
charging LED on the printer flashes orange and green when the Spot LXi is plugged into
the AC main power source. If the Spot LXi has a mounted, external printer that is not
connected via the cable, the charging LED on the printer will not flash.
A dead battery may result if the Spot LXi is left uncharged or shipped/stored for a long
period of a time. If this occurs use the supplied transformer to plug the Spot LXi into the
AC line. Do not use the device for two hours. In extreme cases, the charge condition LED
does not blink (to indicate a fast charge) or the device loses the time and date stamp. If
this happens unplug the accessories and plug the Spot LXi into the AC line using the
appropriate adapter. If the Spot LXi still fails to indicate a fast charge, unplug the AC
adapter, disconnect and reconnect the battery, and plug in the AC adapter. Leave unused
for two hours.
Power On/Off
Press the Power button to turn the device on or off. Upon each power up, the display
lights up, a beep sounds, and the Spot LXi displays the model and serial numbers. If the
internal self-check is successful, the display shows its normal functions (see Figure 4 on
page 24) with all values blank, and the device is ready for operation. If the self-check fails,
an error code is shown in the Navigation Window.
Spot LXi automatically powers off when not used for 30 minutes.
Standby Mode
The Standby Mode conserves battery power. The device goes into Standby Mode if it is
not used for two minutes. Press any button to bring the Spot LXi out of Standby Mode.
Note
There is no hazard associated with leaving the battery in the device, even if the
device is not used for long periods of time.