Product Manual

8 Introduction Welch Allyn Spot Vital Signs LXi
Braun Thermoscan PRO 4000
These warnings are specific to the Braun ThermoScan PRO 4000 thermometer option.
SpO
2
Warnings
These warnings pertain to the Spot LXi SpO
2
feature.
WARNING Keep the probe window clean, dry, and undamaged at all times to
ensure accurate measurements. To protect the probe window, always keep the
thermometer in the storage cover while transporting or when not in use.
WARNING Only use Braun ThermoScan probe covers with this thermometer.
Using other manufacturer’s probe covers or no probe cover may produce
temperature measurement errors and/or inaccuracies. If the thermometer is used
without a probe cover attached, clean the lens (see “Braun ThermoScan PRO
4000 Thermometer” on page 64).
WARNING Do not autoclave.
WARNING The thermometer is not waterproof. Do not immerse or drip fluids
on it. Should this occur, dry the thermometer with warm air. Check for proper
operation and accuracy.
WARNING Only use Spot LXi with Masimo or Nellcor SpO
2
option with Masimo
or Nellcor brand sensors and accessories, respectively. Using the wrong or
unapproved sensors or cables may cause improper performance.
WARNING The SpO
2
sensor and extension cables are intended for use only for
pulse oximetry measurements. Do not attempt to connect these cables to a PC
or any similar device.
WARNING Before using, carefully read the sensor Directions for Use, including
all warnings, cautions, and instructions.
WARNING Do not use a damaged sensor or pulse oximetry cable or a sensor
with exposed optical components.
WARNING Incorrect application or a long duration of use of an SpO
2
sensor may
cause tissue damage. Inspect the sensor site periodically as directed in the
sensor’s Directions for Use.
WARNING Certain ambient environmental conditions, sensor application errors,
and certain patient conditions may affect SpO
2
readings and pulse signal.
WARNING Do not immerse the sensor or patient cables in water, solvents, or
cleaning solutions (the sensors and connections are not waterproof). Do not use
irradiation, steam, or ethylene oxide for sterilization.
WARNING The SpO
2
in the Spot LXi device is not intended for use as an apnea
monitor.
WARNING Consider the SpO
2
an early warning device. As a trend toward
patient deoxygenation is indicated, use laboratory instruments to analyze blood
samples to completely understand the patient’s condition.