56 Chapter 4 Alarms and Alerts Welch Allyn Vital Signs Monitor 300 Series
6. Change or accept the MAP high alarm limit as described above (from step 2).
7. Press to step to the MAP low alarm limit.
8. Change or accept the MAP low alarm limit.
The range of high and low alarm limits for each vital sign is shown here:
Alarm Limits for Vital Signs
Vital Sign Low Limit (Default) High Limit (Default) Resolution
Systolic (mmHg)
Neonatal 30 - 110 (50) 35 - 115 (100) 5 mmHg (0.5 kPa)
Pediatric 35 - 150 (75) 40 - 155 (145) 5 mmHg (0.5 kPa)
Adult 35 - 250 (75) 40 - 255 (220) 5 mmHg (0.5 kPa)
Diastolic (mmHg)
Neonatal 15 - 95 (30) 20 - 100 (70) 5 mmHg (0.5 kPa)
Pediatric 20 - 120 (35) 25 - 125 (100) 5 mmHg (0.5 kPa)
Adult 25 - 225 (35) 30 - 230 (110) 5 mmHg (0.5 kPa)
Mean Arterial Pressure (mmHg)
Neonatal 15 - 100 (35) 20 - 105 (80) 5 mmHg (0.5 kPa)
Pediatric 20 - 130 (50) 25 - 135 (110) 5 mmHg (0.5 kPa)
Adult 25 - 245 (50) 30 - 250 (120) 5 mmHg (0.5 kPa)
Pulse Rate (beats per minute)
Neonatal 35 - 210 (100) 40 - 215 (200) 5 bpm
Pediatric 35- 210 (50) 40 - 215 (150) 5 bpm
Adult 35 - 210 (50) 40 - 215 (120) 5 bpm
SpO
2
(%)
Neonatal 50 - 98 (85) 51 - 99 (95) 1%
Pediatric 50 - 98 (85) 51 - 99 (---) 1%
Adult 50 - 98 (85) 51 - 99 (---) 1%