User Manual User guide
Reference Guide Patient monitoring 59
Apnea events
In the Adult and Pediatric Mode, you can set the apnea delay to 6, 10, 15, 20, 25, or 30
seconds. In the Neonatal Mode, you can set the apnea delay to 6, 10, 15, or 20 seconds.
The Propaq initiates an alarm in response to each apnea event longer than the apnea delay
setting.
When an apnea event is detected, the BR numeric automatically goes to 0 and an apnea
alarm occurs. After the alarm ceases, the Propaq prints an Apnea Ticket if the Apnea
Ticket setting in the Printer Setup window is set to ON.
Numeric area status messages
The following status messages can appear in the numeric display area:
OFF indicates no CO
2
source is selected.
SRCH indicates the MCO
2
or SCO
2
sensor is preparing for a measurement.
UNCAL indicates the monitor has detected a problem such as a lack of calibration, an
obstruction, or a low battery.
WARM UP indicates mainstream CO
2
has been activated and is preparing for
operation. This typically requires 30 seconds at room temperature.
START UP indicates sidestream CO
2
has been activated and is preparing for
operation. This typically requires 30 seconds at room temperature.
CO
2
display menus and status window
To access the first CO
2
menu, press SpO2/CO2, CO2 from the Main Menu
:
Mainstream CO
2
menu and status window
When mainstream CO
2
is active, press MORE from the first CO
2
menu to access the
MCO
2
menu and the MCO
2
status window:
RANGE Selects the CO
2
waveform scale (range)
mm/s Sets the display sweep speed for CO
2
and RESP
RANGE PREV MENUmm/s
MORE
GAS COMP Selects the measurement compensation for CO
2
measurements
RESPONSE Sets the response time for CO
2
measurement (NORMAL, FAST, OR
SLOW)
mmH
g
GAS COMP PREV MENURESPONSE
MCO
2 BR
S
p
O
2
PA
CO
2
GAS COMPENSATION: OFF
RESPONSE : NORMAL
CO2 SOURCE : MAINSTREAM
SWEEP SPEED : 6.25 mm/s
BAROMETER : 762.0 mmHg
35/ 18
( 85)
35
12
92
SOURCE