User Manual User guide
Reference Guide Getting started 25
Printer functions
Press SETUP, MORE, PRINTER to display the printer menu and setup window.
NEXT Selects the next setting in the status window.
CHANGE Changes the currently selected display setting.
PR TREND Prints all trends turned on in the Printer Trend Select Window.
MORE Pressing the MORE button displays another menu and status
window.
PREV MENU Returns you to the previous menu.
CONTINUOUS Sets the print speed for real time (continuous) measurements to
6.25, 12.5, or 25 mm/sec. This sets the print speed for a printout
obtained by pressing the START/STOP button on the printer.
AUTO PRINT Automatically prints 8 seconds of patient information every 15
minutes, 30 minutes, 1 hour, 2 hours, or 4 hours. This is the
latest patient information (real time). The print speed is
automatically set to 25 mm/sec.
ALARM PRINT Automatically prints upon an alarm. The Propaq Encore prints 20
seconds of patient information. The first 12 seconds contain
information prior to the alarm. The print speed is automatically
set to 25 mm/sec.
NIBP TICKET Automatically prints an NIBP Ticket when the measurement is
taken.
APNEA TICKET When turned on, an Apnea Ticket is printed at the conclusion of
an apnea alarm and at the one-minute clock interval if the apnea
alarm does not cease.
OXYCRG ON
ALARM
When turned on, an oxycardiorespirogram will print if an HR/PR,
SpO
2
, or RR/BR alarm occurs. For more information on OxyCRG,
see “OxyCRG” on page 82.
mmH
g
NEXT MORE PREV MENUPR TRENDCHANGE
MCO
2 BR
S
p
O
2
PA
ART
HRT
1NIBP
18:45:28
III
1
mV
cm
MON
PRINTER SETUP PAGE
CONTINUOUS : 25.0 mm/s
AUTO PRINT : 15 minutes
ALARM PRINT : OFF
NIBP TICKET
: OFF
APNEA TICKET : OFF
OXYCRG ON ALARM: OFF