Reference Guide Specifications 119
P1, P2 Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
P1, P2 Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
SpO
2
Limits Adult: 85%, 100%
Ped: 85%, 100%
Neonate: 80%, 95%
RR/BR Adult: 5, 30 BrM
Ped: 10, 45 BrM
Neonate: 10, 60 BrM
TEMP Limits 35.0°, 37.8° C
ΔT Limits 0.0°, 2.8° C
ETCO
2
Limits 25, 60 mmHg (3, 8 for % and kPa)
INCO
2
Limits N/A, 5 mmHg (0.7 for % and kPa)
Apnea Delay Adult/Ped: 20 seconds
Neonate: 15 seconds
PRINTER SETTINGS
Printer Alarm Print OFF
Printer Auto Print OFF
Printer NIBP Ticket OFF
Printer Apnea Ticket ON
Printer Print Speed 25 mm/sec
Printer Auto Trend OFF
Printer Trend Selections NIBP and P1 = ON; all others = OFF
Printer OxyCRG on Alarm OFF
a. Any time you change the Date, Filter, Temp F/C, Decimal, or CO
2
Units setting, the new setting also becomes the powerup default
setting.
Characteristic Specification