User Manual User guide
118 Specifications Welch Allyn Propaq Encore Vital Signs Monitor
Factory default settings
Characteristic Specification
Date
a
MO/DA/YR, DA.MO.YR, or YR/MO/DA
Decimal
a
. (Period)
HR/PR Sweep 25 mm/sec
RR/BR Sweep 6.25 mm/sec
Alarm Tone MEDIUM
HR/PR TONE LOW
HR/PR SOURCE ECG
RR/BR Source CO
2
if ON or ECG (if CO
2
OFF)
Patient Mode Adult
ECG Bandwidth Monitor
ECG Size 1 mV/cm
ECG Lead II
ECG Filter
a
60 Hz
ECG Pacer ON
RESP size 2X
RESP lead Ld2
RESP sweep 6.25 mm/sec
RESP ON/OFF ON
RESP window ON
IBP Range 0 to 180 mmHg
IBP Rescale 0 to 140 mmHg
IBP Mode RESCALE
Invasive Pressure Formats Label dependent
NIBP Mode MANUAL
NIBP Auto Time 15 min
SpO
2
SIZE 2x
SpO
2
C-LOCK OFF
SpO
2
Response NORMAL
TEMP F/C
a
Celsius
CO
2
Range 0 to 60 mmHg
CO
2
Sweep 6.25 mm/sec
CO
2
Response NORMAL
CO
2
Units
a
mmHg
CO
2
Gas Compensation OFF
Sidestream CO
2
Flow Rate Adult: 90 ml/minute
Ped: 90 ml/minute
Neonate: 90 ml/minute
(The flow rate cannot be programmed to a different value in a Custom Patient Mode.
See Custom Patient Modes.)
Display Wave Select All waves are on except NIBP
Trend Group NIBP
Alarm Limits All are ON except P2
HR Limits Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute
NIBP Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg
NIBP Limits - Diastolic Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg
NIBP Limits - Mean Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg
P1, P2 Limits - Systolic Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg