User Manual User guide

114 Specifications Welch Allyn Propaq Encore Vital Signs Monitor
Monitor (physical)
Characteristic Specification
Protection classifications, all configurations
a
a. Per EN 60601-1 unless otherwise stated.
Type of Protection against Electric Shock:—Power
Adapter
Power adapter class 1
Type of Protection against Electric Shock—Monitor
(when connected to power adapter or powered by
internal battery)
Protective earth not available in monitor. Monitor
designed and tested to meet Double Insulation
Requirement.
Degree of Protection Against Electric Shock, for Parts
Applied to Patients
See monitor labels
Recovery time following defibrillator discharge Less than or equal to 10 seconds
Electrosurgery interference suppression Suitable for use in the presence of electrosurgery
Method of Disinfection Not suitable for autoclaving
b
b. See Chapter 7 for cleaning instructions.
Flammable Anesthetics Not suitable for use with flammable anesthetics
Monitor only
Height 6.65 in (16.9 cm)
Width 8.25 in (20.9 cm
Depth 5.10 in (12.9 cm)
Weight 6.25 lb (2.8 kg)
Monitor with SpO
2
module
Height 6.65 in (16.9 cm)
Width 8.25 in (20.9 cm)
Depth 7.50 in (19.10 cm)
Weight 9.12 lb (4.10 kg)
Monitor with expansion module
(Printer / SpO
2
/ CO
2
)
Height 9.65 in (24.5 cm)
Width 8.25 in (20.9 cm)
Depth 7.56 in (19.2 cm)
Weight with Printer, SpO
2
, and CO
2
13.5 lb (6.1 kg)