User Manual User guide
Reference Guide Specifications 105
Temperature
Characteristic Specification
Range 0° to +50°C; 32° to +122°F
Displays T1, T2, and
ΔT
Probes Compatible with YSI Series 400 and 700 probes. HP side panel
only compatible with YSI 400 and has HP connector.
Units °C and °F selectable
Channel Accuracy Temperature Range Tolerance
0° to +10°C
>10° to +50°C
+32° to +50°F
>50° to +122°F
± 0.2°C
± 0.1°C
± 0.4°F
± 0.2°F
Resolution 0.1°C or °F
Temperature Alarm
Limit Range (T1, T2)
32.0° to 122.0°F
0° to +50.0°C
Electrosurgery
interference
suppression
Included in all models: 202EL, 204EL, 206EL