Reference Guide Specifications 103
Invasive pressure
Applies only to models 204 and 206.
Characteristic Specification
Transducer Type Strain-gauge resistive bridge, or HP quartz (with HP Option).
a
a. Transducers with 40 µV/V/mmHg sensitivity are not compatible.
Transducer Excitation Impedance
Range
200 to 2000 Ω
Transducer sensitivity 5 μV/V/mmHg
Excitation Voltage 4.85V Pulsed dc @ 181 Hz
b
b. Duty factor depends on transducer impedance. For 200 to ~900 Ω, duty factor is ≈ 11%. Above ~900 Ω, the duty
factor increases to ≈ 91%.
Connector ITT-Cannon plug MS3106F-14S-6P Std.
Hewlett-Packard compatible 12-pin connector (optional).
Bandwidth Digital filtered, dc to 20 Hz
Zero Drift ± 1 mmHg without transducer drift
Zero Adjustment ± 200 mmHg including transducer offset
Numeric Accuracy ± 2 mmHg or 2% of reading, whichever is greater, plus
transducer error
Pressure range -30 to 300 mmHg
Pulse range 25 to 250 beats per minute
c
c. At pulse rates exceeding 250 beats per minute, refer to the IBP waveforms on the display or printout to
determine systolic and diastolic pressures.
IBP Alarm Limit Ranges All patient modes
Systolic, Diastolic, Mean
-30 to 300 mmHg
Leakage Current Meets ANSI/AAMI risk (leakage) requirements
Electrosurgery interference
suppression
Included in all models
HP 12-pin IBP connector
(side-panel view)
+ EXC
+ EXC
- EXC
- EXC
+ SIGNAL
- SIGNAL
+ SIGNAL
- SIGNAL
SHIELD
SHIELD
Standard 6-pin IBP connector
(side-panel view)