Transducer accuracy test
Required tools, equipment, and accessories
To complete the transducer accuracy test, the following tools and accessories are
required:
scissors or other cutting device
one 4-way hose barb tee for 1/8 inch ID tubing
a minimum of 32 inches of 1/4 inch (.25) OD and 1/8 inch (.125) ID silicone rubber
tubing
approximately 14 inches from the hand bulb to the 4-way tee
approximately 12 inches from the test volume repair fixture to the 4-way tee
approximately 6 inches from pressure meter simulator to the 4-way tee
For the transducer accuracy test, the following equipment is required: (1) the test
volume repair fixture with test manifold, bulb, and valve; (2) a pressure meter simulator;
(3) the device with the Flexiport
®
connector removed. For further information or to order
the test equipment, contact Welch Allyn Technical Support at: http://
www.welchallyn.com/support.
No. Item
1 Test volume repair fixture
(Welch Allyn BP Test Volume Repair Fixture 407672 or equivalent
with test manifold, bulb and valve, 4-way hose barb tee, and 1/4
inch [.25] OD silicone rubber tubing)
Note
For the transducer accuracy test, connect the
test equipment shown below. Connect the
manifold to the test volume repair fixture. The
calibrated volume is 500 ml.
2 Pressure meter simulator (with an accuracy of ± 3 mmHg)
3 Device
(ProBP™ 2000 Digital BP Device)
1. Use scissors or other cutting device to cut off the Flexiport
®
hose fitting from the
end of the device tubing.
Directions for use Specifications 19