User Manual Manual

66 Monitoring and Measurements Welch Allyn 1500 Patient Monitor
NIBP settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description
NIBP Interval Off, 3, 5, 10, 15, 30, 60
Format Sys/Dia or Mean. This defines the main measurement to be
displayed, that is the larger measurement displayed in the NIBP box.
The secondary measurement is displayed smaller by the side of the
defined measurement. Note that the mean measurement is shown in
brackets.
Initial Inflation Pressure Adult: 160
Pediatric: 120
Neonate: 90
NIBPs (systolic blood pressure)
upper/lower limit
Range: 30 - 300
Adult 220/75
Pediatric: 145/75
Neonate: 100/50
NIBPd (diastolic blood pressure)
upper/lower limit
Range: 20 - 235
Adult 110/35
Pediatric: 100/35
Neonate: 70/30
NIBPm (mean pressure) upper/
lower limit
Range: 20 - 255
Adult 120/50
Pediatric: 110/50
Neonate: 80/35