28 Operation Welch Allyn 1500 Patient Monitor
Waveform display
The waveforms that can be configured are as follows:
• RESP in waveform 2 can only be displayed when etCO
2
is not enabled.
• When ST is enabled SpO
2
is available for waveform 4 only. When ST is not enabled
SpO
2
is available for waveform 3 only.
• When resting ECG is enabled, ECG waveform can also be set in waveform 4.
• Waveforms P1, P2, P3, P4 are displayed when selected in the parameter
measurements fields and shown in waveform fields 4, 5, 6 and 7 or 5, 6, 7 and 8 only.
• The size of the waveforms are automatically adjusted for the number of waveforms
displayed.
Waveform
position
ECG RESP CO2 Trend ST SpO2 P1 P2 P3 P4
1 •
2 •••••
3 •••• •
4 ••••
5 ••
6 ••
7 ••
8 •
Waveform 1 - ECG (I, II, III, V, aVL,
aVR, aVF)
Waveform 2 - ECG, RESP, CO
2
, ST
or Trend (“Trend data” on page 29)
Waveform 3 - ECG, RESP, Trend,
CO
2
, or SpO2
(Waveform 4- ECG, Trend, or SpO
2
)
Waveform 4 to 7 (5 to 8) - P1 to P4,
or P1 and P2, or no IBP waveform
displayed. P1, P2, P3, P4 will be
displayed as selected for
parameter measurement (see
“Defining parameter fields” on
page 25.)