132 Technical data Welch Allyn 1500 Patient Monitor
Respiration
Temperature
ECG amplifier
Sampling frequency 1000 Hz
Pacemaker detection ± 2 to ± 700 mV / 0.1 to 2 ms
Pacemaker rejection ± 2 to ± 700 mV / 0.1 to 2 ms
Note: Pacemaker signals can differ from one pacemaker to the next. Rate
meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias, mainly with pacemakers generating
high amplitudes ( > 20 mV) or those generating overshoot. Pacemaker
patients should be kept under close or constant observation.
Protection Fully isolated, defibrillation protected >5 kV
Line frequency filter 50 or 60 Hz sinusoidal interferences filtered by means of adaptive digital
filtering.
Respiration rate range 0 to 200 breaths / min (pediatric: 0 to 120 breaths per minute)
Connector Shared with ECG
Signal 28 kHz square wave ± 2.5 V
Patient current max. 80 :A
Dynamic impedance
range
1k to1.5 kS, variation of 0.1 to 3 S
Sampling Rate 250 Hz
Respiration rate accuracy ± 1 digit
RR display update interval Max. 2 s
Channels One or two channels
Sensor YSI 401, rectal, skin or ear
Amplifier Fully isolated, defibrillation protected >5kV
Sampling Frequency 125 Hz
Measurement interval 1x per second
Measurement range 15 °C to 45 °C (59 °F to 113°F)
Resolution 0.1 °C (0.1 °F)
Accuracy +
0.1° C (+ 0.1° F)