User Manual Manual
128 Technical data Welch Allyn 1500 Patient Monitor
Printer High-resolution thermal printer
Resolution 8 dots/mm (amplitude-axis), 40 dots/mm (time-axis) at 25 mm/s
Paper Thermoreactive, Z-folded
Width: 80 mm
Length 20 m (approx.)
Print speed 25 mm/s
Printout length 10 second ECG recording on 4 pages
Recording tracks 3-channel display, with optimal width of 72 mm, automatic baseline
adjustment
Printout Curves, trend and saved values
Battery
Battery type Lead acid battery, 12 V
Capacity 2600 mAh
Recharging time 80% capacity: 2.8 hours
100% capacity: 3.5 hours
(monitor switched off)
Battery life up to 1000 cycles
or
Battery type Lithium-Ion battery, 10.8V
Capacity 7200 mAh
Recharging time 80% capacity: 2.5 hours
100% capacity: 6.5 hours
(monitor switched off)
Battery life min. 500 cycles
Connections ECG
SpO
2
NIBP
etCO
2
CO
Temperature (x1) or (x2)
Invasive blood pressure (x2) or (x4)
Interfaces Ethernet via RJ45
Nurse call:
Alarm delay at the signal output component <0.5 s
Plug type: 1/8 in (3.5 mm) mini-phone jack stereo connector
Tip: Normally closed
Ring: Normally open
Maximum switch current: 1A
Maximum switch voltage: 30 V AC/DC
Isolation: 1,000 Vrms for 1 min
USB 1.1
Demo Mode Simulated patient information including waveforms for training and
education