Neonatal settings
Parameter Acuity range limit Acuity lower limit
(default)
Acuity upper limit
(default)
HR 25 - 250 100 200
PR (NIBP) 25 - 250 100 200
PR (IBP) 25 - 250 100 200
PR (SpO
2
) 25 - 250 100 200
SpO
2
SAT (%) 50 - 100 85 95
RR-ECG 3 - 150 10 75
RR-CO
2
3 - 150 10 75
Apnea Delay (s) 6 - 20 N/A 15
etCO
2
(mmhg) 0 - 99 25 60
etCO
2
(kPa) 0 - 13.2 3.0 8.0
inCO
2
(mmhg) 2 - 25 N/A 5
inCO
2
(kPa) 0.2 - 5.0 N/A 0.7
P1 Sys -30 - 300 50 100
P1 Dia -30 - 300 30 70
P1 Mean -30 - 300 35 80
P2 Sys -30 - 300 15 50
P2 Dia -30 - 300 5 20
P2 Mean -30 - 300 10 25
P3 Sys
P3 Dia
P3 Mean
P4 Sys
P4 Dia
P4 Mean
NIBP Sys 25 - 120 50 100
NIBP Dia 15 - 105 30 70
NIBP Mean 15 - 110 35 80
Temp (F) 59 - 113 95.0 100.0
Temp (C) 15 - 45 35.0 37.8
Temp2 (F)
Temp2 (C)
Temp (F)
Temp (C)