96 Acuity Central Station Welch Allyn 1500 Patient Monitor
Pediatric settings
Parameter Acuity range limit Acuity lower limit
(default)
Acuity upper limit
(default)
HR 25 - 250 50 150
PR (NIBP) 25 - 250 50 150
PR (IBP) 25 - 250 50 150
PR (SpO
2
) 25 - 250 50 150
SpO
2
SAT (%) 50 - 100 90 100
RR-ECG 2 - 150 10 45
RR-CO
2
2 - 150 10 45
Apnea Delay (s) 6 - 30 N/A 20
etCO
2
(mmhg) 0 - 99 25 60
etCO
2
(kPa) 0 - 13.2 3.0 8.0
inCO
2
(mmhg) 2 - 25 N/A 5
inCO
2
(kPa) 0.2 - 5.0 N/A 0.7
P1 Sys -30 - 300 75 145
P1 Dia -30 - 300 35 100
P1 Mean -30 - 300 50 110
P2 Sys -30 - 300 15 50
P2 Dia -30 - 300 5 20
P2 Mean -30 - 300 10 25
P3 Sys
P3 Dia
P3 Mean
P4 Sys
P4 Dia
P4 Mean
NIBP Sys 30 - 260 75 220
NIBP Dia 20 - 235 35 110
NIBP Mean 20 - 255 50 120
Temp (F) 59 - 113 95.0 100.0
Temp (C) 15 - 45 35.0 37.8
Temp2 (F)
Temp2 (C)
Temp (F)
Temp (C)