settings
The parameter settings are selected with the trim knob, described previously (see
“Settings via a parameter field” on page 31).
The default settings are in bold.
Parameter Description
SpO
2
alarm pause When the alarm is paused the message SpO
2
Alarms Paused
appears in the message line. The alarm is paused until the sensor is
reinstalled.
HR/PR source
1
1. This can also be set in the Setup HR menu.
ECG, SpO
2
, P1 (selection of the heart rate source). The source is
displayed in brackets next to the HR parameter.
HR/PR tone
1
Off, on
2
2. If SpO
2
is the HR/PR source, the pitch of the beep corresponds to the SpO
2
saturation. A high-pitched beep
indicates a high saturation.
SpO
2
lower/upper limit Range: 50- 100
Lower limit default 50
Upper limit default 100