User Manual
Directions for use Settings 103
Recorder
This defines the information to be printed.
The waveforms that can be printed will depend on the options enabled and the
configuration of the monitor. The default settings are in bold.
* These parameters are available only when installed and enabled.
** This option is not available when CO
2
is enabled.
Parameters
The parameters that can be enabled and displayed are set in this menu. The parameter
options will change according to the licensed options and monitor configuration and will
be any combination of the following:
•etCO
2
•ST
• Masimo settings (extra settings for SpO
2
)
• CO (cardiac output)
• IBP channel (number of invasive BP channels (none, 2 or 4)
• Temperature display mode (T1, T1 and T2, or T1 and T)
Details of the displayed parametrize are given in the operation section (see “Defining
parameter fields” on page 25).
Note
This is only available when the recorder option is installed.
Main menu Parameter Description
SetUp Recorder Waveform 1 ECG1, ECG2, ECG3, RESP, CO
2
*, SpO
2
, P1, P2, P3*, P4*
Waveform 2 ECG1, ECG2, ECG3, RESP**, CO2 *, SpO
2
, P1, P2, P3*,
P4*, OFF
Waveform 3 ECG1, ECG2, ECG3, RESP, CO
2
*, SpO
2
, P1, P2, OFF
Recording time 5, 10, 16 Seconds
Recording Delay 0, 6, 10 Seconds. This defines the duration of data that is
printed before the print key is pressed. For example, if a
delay of six seconds with a recording time of 10 seconds
is defined (default), six seconds of data recorded before
the print key is pressed is printed, followed by the
subsequent 4 seconds.