Owner's manual
28
Set voltage source to 10 V. Note: the error message SHrt indicates a bad connection.
U=10
Off Off On
Ó
Calibrate 10 V point
Sure Ó
To Save Values (or any other key to restart).
External Current Inputs
Display I+ I- V Keys Action
Switch on Power, Calibrate link In CAL Position.
u=0
Off Off On
Ô Õ Ù Ó
Press sequence to select External Voltage
mode. INC and DEC Leds will both light.
Connect current source and Set to 0 mA.
I=0
On Off Off
Ó
Calibrate 0 mA point
Set current source to 20 mA. Note: the error message Open indicates a bad connection.
I=20
On Off Off
Ó
Calibrate 20 mA point
Sure Ó
To Save Values (or any other key to restart).
External Voltage Inputs (continued)