Datasheet
Data sheet
MCZ SERIES
MCZ PT100/3 CLP -50C...+150C
Weidmüller Interface GmbH & Co. KG
Klingenbergstraße 26
D-32758 Detmold
Germany
Fon: +49 5231 14-0
Fax: +49 5231 14-292083
www.weidmueller.com
Technical data
Creation date January 24, 2020 3:25:37 PM CET
Catalogue status 13.01.2020 / We reserve the right to make technical changes. 2
Dimensions and weights
Length 91 mm
Length (inches) 3.583 inch
Width 6 mm
Width (inches) 0.236 inch
Depth 63.2 mm
Depth (inches)
2.488 inch
Net weight 29.2 g
Temperatures
Operating temperature, max. 50 °C
Operating temperature, min. -25 °C
Storage temperature, max. 85 °C
Storage temperature, min. -25 °C
Operating temperature -25 °C...50 °C
Storage temperature -25 °C...85 °C
Probability of failure
MTTF 1,068 Years
Environmental Product Compliance
REACH SVHC Lead 7439-92-1
Rated data UL
Certificate No. (cURus) E141197
Input
Number of inputs
1
Sensor PT100/2-/3-wire (in
compliance with IEC 751)
Sensor supply 0.8 mA / 9…30 V DC
Temperature input range -50...+150 °C
Output
Number of outputs
1
Output current 4...20 mA (current loop) at
9...30V DC
load impedance current ≤ 600 Ω
General data
Accuracy Typ.: 0.2 %, máx. 0.5 % v.
FSR
Configuration
none
Galvanic isolation
between input/output
Input/Output PT100/ 4...20 mA (current
loop)
Linearity Yes
Rail TS 35
Step response time 10 ms
Temperature coefficient
max. ± 250 ppm/K
Voltage supply Output loop powered,
max. 30 V/ min. 9 V + (20
mA x RL)
Insulation coordination
EMC standards EN 61000-6
Galvanic isolation between input/output
Pollution severity 2
Rated voltage
50 V
Surge voltage category I