User Documentation
232661740000/01/06-2020
User commissioning
User commissioning
Step 1 Touch panel calibration:
The user is requested to ‘touch’ using the stylus on the top right-hand corner of the frame, and
then again on the bottom left-hand corner. If the touch is within expected limits, the coordinates are
accepted and the display calibrated accordingly.
Step 2 Instrument self-calibration:
Please follow the instructions to rst remove the test leads from the front terminal sockets and then
press the Continue... button.
The red high voltage LED will briey illuminate, indicating that high voltage is being produced at
the terminals. Internal calibration is now complete.
Step 3 Observe the Splash screen warning – clicking the Acknowledge... button to signify that the warn-
ing has been read and the potential hazard is understood.
Step 4 Enter the default passcode 1_2_3_4 to unlock the instrument.
The instrument enters the Main screen and the blue LED extinguishes. The user can now proceed to
customize the instrument in the Device Setup and Measure Setup menus.
Device setup
From the Main screen navigate to: SETUP / Device Setup
The DEVICE SETUP MENU, allows the user to select one of the following sub-menus:
• Set Time/Date
• Set Passcode
• Set Display
• Set Options
Click the < BACK button to return to the Main screen.
Measure setup
From the Main screen navigate to: SETUP / Measure Setup
The Measurement Settings... menu, allows the user to select one of the following sub-menus:
• set GDT Ramp Rate
• set MOV Constant Current
Click the < BACK button to return to the Main screen (refer to Measurement screen).