Table 2. Guidance and manufacturers declaration – electromagnetic immunity
The TS200 is intended for use in the electromagnetic environment specified below.
The customer or the user of the TS200 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
(50, 60 Hz) magnetic
field
IEC 61000-4-8
IEC 60601 test level
5
.
E
l
e
c
t
r
o
m
a
g
n
e
t
i
c
c
o
m
p
a
t
i
b
i
l
i
t
y
i
n
f
o
r
m
a
t
i
o
n
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Compliance level
Electromagnetic environment – guidance
Contact:±8 kV
Air:±2 kV,±4 kV,±8 kV,
±15 kV
+ 2kV for power supply
lines
+ 1kV for input/output
lines
+ 0.5kV, +1kV line(s)
to line(s)
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30 cycles
Voltage interruptions:
0 % UT; 250/300 cycle
Voltage dips:
Not applicable
Not applicable
Not applicable
Voltage interruptions:
Not applicable
Contact:±8 kV
Air:±2 kV,±4 kV,
±8 kV,±15 kV
NOTE
U
T is the a.c. mains voltage prior to application of the test level.
The TS200 power frequency magnetic
fields should be at levels characteristic
of a typical location in a typical home
healthcare environment.
Mains power quality should
be that of a typical home
healthcare environment.
Mains power quality should
be that of a typical home
healthcare environment.
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Mains power quality should be
that of a typical home healthcare
environment. If the user of the
TS200 requires continued operation
during power mains interruptions,
it is recommended that the TS200
be powered from an uninterruptible
power supply or a battery.