Specifications
place the sample holder in the lamp and light it (Warning—Do
not allow the flame to come closer than
1
⁄
8
in. (3.2 mm) from
the thermocouple or the potentiometer (digital indicator-Serial
Nos. 378 or higher) to read 1000°F (538°C) or higher). If any
soot builds upon the thermocouple shield, put out the flame,
clean the shield, lower the sample holder slightly and relight.
Through the observation peep hole in the center of the lamp
door note that the flame is burning free of smoke (Note 7). Let
the tetralin burn at this level for 15 min to warm up the
apparatus.
NOTE 7—All ASTM-CRC Luminometers have been adjusted by the
manufacturer to read 45 to 55 at the smoke point of tetralin. If the unit is
outside this range, adjust the internal resistor located between the two
amplifier tubes (remove the rear cover of the unit) until a meter reading of
45 is obtained.
9.2 After warmup, lower the sample holder until the lumi-
nometer meter indicates approximately 30. Rezero the lumi-
nometer meter and potentiometer (6.1.3). Turn the luminom-
eter meter switch to the TEST position. Wait at least 30 s for
the indicator to stabilize after switching. Operate the luminom-
eter at least 5 min at this position with the outer door closed in
a draft-free environment. Record the luminometer meter and
temperature indications (Note 8). Raise the wick and its holder
until a luminometer reading approximately five units higher is
indicated. Allow 5 min and again record the luminometer meter
and temperature indications as above. Repeat this procedure
until four data points are obtained; the last one being obtained
at the flame height where a luminous tail (trace smoke) just
breaks out of the tip of the flame. Soot would accumulate on
the thermocouple at a slightly higher setting.
9.3 Plot the data points on a luminometer meter reading
versus lamp temperature rise curve. All points should fall on a
smooth curve. The top point (smoke point of tetralin) will
represent rating level (meter reading) for all samples to be
tested in this instrument. Repeat four times to establish the
average rating level for the instrument (see Fig. 2).
NOTE 8—The rebalancing can be facilitated by putting a 40 to 60-mesh
screen in front of the opening.
9.4 Run two samples of ASTM reference fuel grade iso
octane as described in 9.1-9.3, but obtain four data points, two
below and two above the rating level established with tetralin
for the instrument. (The four data points should be uniformly
separated by approximately 10 luminosity units.) One of the
samples should be run before running the unknown test fuel;
the other sample should be run after running the test fuel. Plot
two curves, and at the rating level, find the lamp temperature
rise for each sample of iso octane (Fig. 3). Average these
values.
9.5 Run one sample of the unknown test fuel in the manner
described for iso octane and determine the lamp temperature
rise for this fuel (Fig. 4). Use this value and the average DT’s
of tetralin and isooctane to calculate the luminometer number
of this test fuel.
NOTE 9—If a number of unknown test fuels are to be tested, the second
sample of isooctane may be run after all of the unknown test fuels,
provided all of these runs are made the same day.
N
OTE 10—Fuels that have high luminometer numbers are prone to form
condensation and are greatly affected by air leaks or leveling errors.
Preheating the interior of the cabinet with a small drop light, or using a
slow step-type warmup will prevent condensation problems. If condensa-
tion is encountered, meter readings will become unstable and moisture
will appear on the peep-hole glass. If condensation is encountered, it can
be removed by allowing the fuel to burn with the lamp door removed until
all moisture clears from the filter glass and holder.
N
OTE 11—Fuels that have very low luminometer numbers frequently
smoke at relative low meter readings. The absence of a smoking flame
with fuels giving a low-temperature rise should be assured by observation
Fuel Tetralin
Luminometer No. 02L-1
Barometer, in. Hg 29.92
Wet bulb temperature 59°F (15°C)
Dry bulb temperature 77°F (25°C)
Background to be measured with sample holder
in lamp, no flame, door closed, and Luminometer
meter zeroed
0
Lamp temperature rise at rating level 233°F
FIG. 2 Rating Level of Tetralin
Fuel
Iso
octane
Barometer, in. Hg 29.92
Wet bulb temperature 59°F (15°C)
Dry bulb temperature 77°F (25°C)
Background to be measured with sample holder
in lamp, no flame, door closed, and Luminometer
meter zeroed
0
Lamp temperature rise at rating level 484°F
FIG. 3 Lamp Temperature Rise at Rating Level of
Iso
octane
D1740–01
4







