Specifications

page | 48
• 1 mS/cm = 1000 µS/cm (µS: microSiemens / mS: miliSiemens) • 1 ppt = 1000 ppm (ppm: parts per million / ppt: parts per thousand) ** Maximum 199.9 ppt depending on factor setting
* RS232C (LED) interface adapter available as separate accessory (see page 49 for order information)
Models
CyberScan Dual-Display Eutech Single-Display
COND 610 COND 600 CON 400 CON 110 CON 11 COND 6+ TDS 6+ Salt 6+
Conductivity/
TDS/Salinity
Handheld Meters
Specications
Measuring Parameter
Conductivity / TDS /
Salinity / Resistivity / ºC/ ºF
Conductivity / TDS / ºC / ºF
Conductivity / TDS /
ºC / ºF
Conductivity /
TDS / ºC
Conductivity / ºC TDS / ºC Salinity / ºC / ºF
Highlights
Waterproof, GLP,
RS232C, IrDA,
linear & pure TC
Waterproof, GLP,
RS232C, IrDA,
linear TC
Waterproof, GLP
Expanded
memory, RS232C
output
Standard
handheld
Economical CON
measurement
Economical TDS
measurement
High accuracy over
a wide Salinity
range
Conductivity
Range
.. to 2.000 µS
2.000 to 300.0 µS
300.0 µS to 4.000 mS
4.000 to 40.00 mS
40.00 to 500.0 mS
.. to 2.000 µS
2.000 to 300.0 µS
300.0 µS to 4.000 mS
4.000 to 40.00 mS
40.00 to 200.0 mS
.. to 19.99 µS
19.9 to 199.9 µS
199 to 1999 µS
2.00 to 19.99 mS
20.0 to 199.9 mS
Resolution
0.01 µS / 0.1 µS / 0.001 mS /
0.01 mS / 0.1 mS
0.05 % full scale
Accuracy
±1 % full scale + 1 LSD
TDS
Range
(Depending
On TDS Factor)
.. to 2.000 ppm
2.000 to 300.0 ppm
300.0 ppm to 4.000 ppt
4.000 to 40.00 ppt
40.00 to 500.0 ppt
.. to 2.000 ppm
2.000 to 300.0 ppm
300.0 ppm to 4.000 ppt
4.000 to 40.00 ppt
40.00 to 200.0 ppt
.. to 10.00 ppm **
10.0 to 100.0 ppm
100 to 1000 ppm
1.00 to 10.00 ppt
10.0 to 100.00 ppt
.. to 9.99 ppm **
10.0 to 99.9 ppm
100 to 999 ppm
1.00 to 9.99 ppt
10.0 to 99.9 ppt
.. to 9.99 ppm **
10.0 to 99.9 ppm
100 to 999 ppm
1.00 to 9.99 ppt
10.0 to 99.9 ppt
Resolution
0.01 ppm / 0.1 ppm / 0.001 ppt /
0.01 ppt / 0.1 ppt
0.05 % full scale 0.05 % full scale
Accuracy
±1 % full scale + 1 LSD ±1 % full scale + 1 LSD
TDS Factor
0.40 to 1.00 0.40 to 1.0 0.40 to 1.0 0.40 to 1.0
Salinity
Range
.. to 0.770 ppm
0.770 to 143.3 ppm
143.3 ppm to 2.138 ppt
2.138 to 23.64 ppt
23.64 to 80.00 ppt
1.0 to 50.0 ppt / 0.1
to 5.00 %
Resolution
0.01 ppm / 0.1 ppm /
0.001 ppt / 0.01 ppt
0.1 ppt / 0.01 %
Accuracy
±1 % full scale + 1 LSD
±1 % full scale
Resistivity
Range
2.000 to 25.00 Ω
25.00 to 250.0 Ω
250.0 to 3.333 kΩ
3.333 to 500.0 kΩ
500.0 kΩ to 20.00 MΩ
Resolution
0.01 Ω / 0.1 Ω / 0.001 Ω /
0.1 Ω / 0.01 MΩ
Accuracy
1 % full scale + 1 LSD
Cal. Points
4 (1 per range) auto, 5 (1 per range) manual 5 (1 per range) manual 4 (1 per range) auto, 5 (1 per range) manual 5 (1 per range) manual
Temperature
Range
-10.0 to 110 ºC / 14.0 to 230 ºF 0.0 to 100.0 ºC / 32.0 to 212 ºF -10.0 to 110 ºC
Resolution
0.1 ºC / 0.1 ºF 0.1 ºC
Accuracy
±0.5 ºC / ±0.9 ºF ±0.5 ºC
Compensation
ATC / MTC (0 to 80 ºC) ATC / MTC (0 to 50 ºC)
Normalization
15 to 30 ºC 20.0 or 25.0 ºC
20.0 ºC & 25.0 ºC
(selectable)
Operating Temp.
0 to 50 ºC 0 to 50 ºC
Meter
Features
Temperature
Coefficient
Linear & pure Linear 0.0 to 10.0 % 0.0 to 3.0 % 0.0 to 3.0 %
GLP
Yes
Cal-Due Alarm
Yes
IP67
Yes
Datalogging
Yes Yes
Memory
500 data sets 50 data sets 100 data sets 50 data sets
Cell Constant
0.010 to 10.000 0.1, 1.0, 10.0
LCD Display
Dot-matrix LCD with backlight (5.4 x 7.1 cm) Dual-display LCD (5.8 x 3.3 cm) Single-display LCD (4.5 x 2.3 cm)
Auto-Off
2 to 30 mins after last key pressed 20 mins after last key pressed
Auto Hold
Yes
Input
DC phono sockets, 8-pin connector 6-pin connector DC socket, 6-pin connector BNC, 2.5 mm phono socket
Output
IrDA, RS232C (via LED) * RS232C
Power
4 x 1.5 V ‘AA’ alkaline batteries or
9 V DC adapter, 500 mA
4 x 1.5 V ‘AAA’
alkaline batteries
4 x 1.5 V ‘AAA’ alkaline batteries or
9 V DC adapter, 200 mA
4 x 1.5 V ‘AAA’ alkaline batteries
Battery Life
> 200 hrs > 100 hrs > 200 hrs > 100 hrs
Dimensions
(LxWxH)
; Weight
Meter
18.3 x 9.5 x 5.7 cm ; 460 g 19 x 10 x 6 cm ; 320 g 18 x 9 x 4 cm ; 220 g 15.7 x 8.5 x 4.2 cm ; 255 g
Boxed
40 x 33 x 10 cm ; 2680 g 40 x 33 x 10 cm ; 2100 g 36 x 28 x 8 cm ; 1555 g
Conductivity/
TDS/Salinity
Handheld Meters Specifications
Conductivity / TDS / Salinity