2. Inspect top temperature sensor and radiation shield
• Remove temperature sensor from radiation shield and inspect sensor for visible damage
o Replace any damaged sensors
• Inspect radiation shield for damage, debris, insects, and spider nests
o Replace any damaged shields. Remove any foreign objects from shield and clean shield
• Once radiation shield is clean, replace top temperature sensor
• Ensure radiation shield is securely mounted to mast
3. Inspect wind sensors
• Check wind speed sensor for damage
o Check wind speed sensor for smooth turning operation
o Listen for any grinding noise while speed sensor is turning
• Check wind direction sensor for damage
o Check wind direction sensor for 360° of smooth operation
o Ensure reference mark is facing north
• Ensure sensors are securely mounted to crossarm
• Ensure crossarm is securely mounted to mast
• Ensure crossarm is level
4. Inspect bottom temperature/RH sensor and radiation shield
• Remove temperature/RH sensor from radiation shield and inspect sensor for visible damage
o Replace any damaged sensors
• Inspect radiation shield for damage, debris, insects, and spider nests
o Replace any damaged shields. Remove any foreign objects from shield and clean shield
• Once radiation shield is clean, replace temperature/RH sensor
• Ensure radiation shield is securely mounted to mast
Do not spray insect/spider killer inside radiation shield while temperature sensor is housed inside
shield. Contact with these chemicals could damage temperature sensor.
Do not spray insect/spider killer inside radiation shield while temperature/RH sensor is housed inside
shield. Contact with these chemicals could damage temperature/RH sensor.