Specifications
User Manual 301/302 
segment is deleted, the next downloaded segment will be written on 
top of the deleted one. There is danger that by using the TRAC:DEL 
command often large portions of memory will remain unused. It is, 
therefore, suggested that you periodically clear the entire memory and 
only reload waveforms that you intend to use. 
To partition the memory from the beginning, use the following 
command: 
TRACe:DELete: ALL 
In WaveWiz use the   button.
CAUTION
This command will destroy waveforms that were previously loaded 
i
nto memory. After using this command, waveform segments will 
l
ine-up from address 0 upwards. 
Loading Arbitrary 
Waveforms
There are three ways to load waveforms into the 301/302 these are 
using 1) WaveWiz, 2) LabVIEW driver, or 3) Low-level programming. 
When using WaveWiz, disregard most of this chapter and move on to 
“Using WaveWiz to Generate Arbitrary Waveforms” as this software 
does the work for you. If you write your own application or instrument 
driver, use the following commands to load data into a specific 
memory segment. 
First, define the work area. Define the segment number and its 
associated length. The segment length must be divisible by eight or an 
error will occur. For example, to use segment number 8 with a 
waveform length of 1048 points use this command: 
TRACe:DEFine 8,1048 
Next, make segment 8 the active segment. The active segment must 
be selected because as waveforms are loaded, the Model 301/302 
must be notified as to where to place the data it receives. Select the 
active segment using the following command: 
TRACe:SELect 8 
The next step is to transfer data to the active segment. Data is loaded 
into the Model 301/302 using high-speed binary transfer. A special 
command is defined by IEEE-STD-488.2 for this purpose. High-speed 
binary transfer allows any 8-bit bytes (including extended ASCII code) 
to be transmitted in a message. This command is particularly useful 
for sending large quantities of data. The Model 301/302 uses this 
command to receive waveforms from the controller: 
Configuring the Instrument 