User`s guide
Interpreting Oscilloscope Views (plots)
• Time - Time vs. Voltage including Eye Mask options.
• Summary - Textual display of Oscilloscope measurements.
Time View with Eye Mask Enabled
To make an Eye Mask measurement, connect a data signal to one channel and a bit clock to another channel.
Set the trigger to be the channel with the bit-clock.
Press Signal Analysis, Eye Mask, then select ON in Enable Eye Mask.
Pressing Show Measures will display the number of samples on the display (Compares) and the number of
Violations of the mask regions. This is a toggle, and pressing it again will turn off those values.
Press Zoom to Eye to automatically place the mask in the display. To change the mask position or size, press
Mask Setup. This will allow you to configure the mask.
For other configuration and tool settings, refer to the context sensitive help. The help describes the use of
each setting.
In Time View, resolution affects the time between measurements. Range affects the time over which
measurements are made.
Section 4 - GigaView
©
WAVECREST Corporation 2005
44