User`s guide
Fuse Replacement
The IEC plug-in at the back of the SIA-3000 provides the input connection for the AC power
cord. A small compartment on the IEC plug houses two Wickmann 10AT/230V fuses (5x20mm).
To gain access to the fuses, remove the power to the SIA-3000 and disconnect the power cord.
With a small screwdriver, pry open the fuse compartment on the IEC plug. Remove the fuses
and install the new fuses prior to closing the compartment and reinstalling the power cord.
Cleaning Instructions
Periodic cleaning may be required. Use a clean, lint free cloth to dust outer surfaces after electrical
power has been removed. If needed, dampen cloth with light amounts of water or alcohol.
Intended Use Of Equipment
The SIA-3000 should be used only for its intended purpose as outlined in this manual. To
avoid possible injury, the instrument should not be operated with the top cover or other panels
removed. Refer installation and maintenance to qualified service personnel.
CAUTION: The internal CPU assembly contains a lithium battery. Danger of explosion exists if battery is incorrectly replaced. Replace
only with the same or equivalent type recommended by the manufacturer. Discard used batteries according to the manufacturer’s instructions.
VORSICHT: Explosionsgefahr bei unsachgemäßem Austausch der Batterie. Ersatz nur durch denselben oder einen vom Hersteller
empfohlenen gleichwertigen Typ. Entsorgung gebrauchter Batterien nach Angaben des Herstellers.
ATTENTION: IL Y A DANGER D'EXPLOSION S'IL Y A REMPLACEMENT INCORRECT DE LA BATTERIE. REMPLACER
UNIQUEMENT AVEC UNE BATTERIE CU MĔME TYPE OU D'UN TYPE RECOMMANDÉ PAR LE CONSTRUCTEUR. METTRE AU
RÉBUT LES BATTERIES USAGEES CONFORMÉNENT AUX INSTRUMCTIONS DU FABRICANT.
To avoid explosion, do not operate the SIA-3000 in or near an atmosphere of explosive gases.
This equipment is to be connected to supply mains using the power cord provided with the
product. Transient over voltages are to be less than the limits defined by installation category
II, Annex J of EN61010-1.
©
WAVECREST Corporation 2005
Appendix D
291