User`s guide

Making a Fibre Channel Measurement
Setup Directions
The tool requires a known repeating pattern such as CRPAT, CJTPAT, K28.5, etc.
If the SIA-3000 has the PM50 Pattern Marker option installed, connect the data directly to Channel 1 (IN1).
Note: If the Pattern Marker Option is not installed, you must have a pattern marker signal
supplied externally. The Marker is a pulse that repeats once per pattern. Pattern
Generators typically have a "pattern trigger" or "pattern sync" that will work. With an
externally supplied pattern marker, uncheck the pattern marker box.
From the Main Dialog Bar, press "Data Standards" and "Fibre Channel" to Open the tool.
Verify the channel settings.
Press Arm Number and select Channel 1 and check Pattern Marker.
Section 4 - GigaView
©
WAVECREST Corporation 2005
198