Caution: Before operating ensure the equipment is firmly secured to eliminate the possibility of causing injury
or damage. This function will power the Unit on, Please make sure the drills or any blades of any rotating
devices have been removed prior to any testing and have clearance from body or object. Appliances may
jump off benches and may need to be secured by clamping them, or they may cause serious damage if not
secured properly.
KILL SWITCH
To stop operation press the return/enter key at any time. The unit will power down after 20 seconds
as factory default
Caution: This test turns the Equipment ON. The power / leakage test can handle up to 10 Amp (Resistive)
Load MAX. Exceeding this value will blow the external M205 ceramic fuse. Care should be taken when
testing inductive loads (i.e. AC motors) with large start up currents.
If 10A HBC fuse is blown the appliance will not operate in Power test or Leakage test modes. Replace fuse
only with M20 10A HBC fuses. Any other fuse voids warranty and manufacturers design.
Caution: DO NOT OPEN TnT - Series case as there are no serviceable parts inside and you can void your
Warranty