Instruction manual

Page 22
Chapter 3 – TnT+ Portable Appliance
Part No: WCM-TNT+
Test Instructions
TnT+ Test Function Flow Chart









Menu 2
F1 Power Test
F2 Leakage Test
F3 RCD Test
Menu 2
F1 POWER TEST
Instructions for POWER TEST:
F1 + F3 = POWER RESULTS
Menu 2
RCD TEST
Instructions for RCD TEST for;
Time Test
O’ Phase - F3 + F1 + F1 + F3 = RESULTS
18O’ Phase - F3 + F1 + F2 + F3 = RESULTS
Change Current F3 + F1 + F3 + F3 = SETTING
Ramp Test F3 + F2 + F3 = RESULTS
Menu 2
F2 LEAKAGE TEST
Instructions for LEAKAGE TEST for;
Earthed Devices: F2 + F1 + F3 = RESULTS
D.Ins., Ex-Lead: F2 + F2 + F3 = RESULTS
RCD Protected: F2 + F3 + F3 = RESULTS
APPLIANCE POWER TEST
Continue if Safe?
OK QUIT

Please note: Menu 1 is common to all models [TnT, TnT-el & TnT+]
(r
efer page 14 for more details)
+
POWER RESULTS
240 V 0 W
0.000 A 0 VA
0.00 PF
LEAKAGE TEST
(F1) Earthed
Devices
(F2) D.Ins., Ex
-
Lead
To return to MENU 2
Press RETURN
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 5.0 mA
0.0 mA
Pass ESC
EMERGENCY STOP
ABORT
To return to MENU 2
Press RETURN
LEAKAGE TEST
(F1) Earthed Devices
(F2) D.Ins., Ex-
Lead
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 1.0 mA
0.2mA
Pass ESC
LEAKAGE TEST
(F1) Earthed Devices
(F2) D.Ins., Ex-Lead
(F3) RCD
Protected
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 2.5 mA
1.0mA
Pass ESC
RCD TESTS
(F1) Time Test
(F2) Ramp Test
TIMED TEST 30mA
(F1) 0’ phase
(F2) 180’ phase
(F3) Change Current
RCD MULTIPLYER
(F1) Up 1.0
(F2) Down
(F3) START
RCD TIMED TEST
Time: 4ms
&
TIMED TEST 30mA
(F1) 0’ phase
(F2) 180’ phase
(F3) Change Current
RCD TEST CURRE
NT
(F1) Up 30mA
(F2) Down
(F3) SET
RCD TEST CURRENT
Up 30mA
Down
SAVING SETTINGS
RCD TESTS
(F1) Time Test
(F2) Ramp Test
RCD RAMP TEST
(F3) START
RCD RAMP TEST
TESTING
RCD RAMP TEST
Time: 4ms