34 Watlow Series 935A User’s Manual
Calibrating the 935A
Calibration requires a precision millivolt source with
thermocouple compensation, an adjustable 0-10 volt source,
and a decade resistance box.
[tr50] and [tr00] calibrate the thermocouple span.
[``tc] calibrates the ambient compensation.
[r380] and [r`15] calibrate the RTD span.
0V is required when calibrating [tc00] and [r`15]
for remote set point calibration.
5V is required when calibrating [tc50] and [r380]
for remote set point calibration.
When calibrating, calibrate all points for consistency in results.
Allow the unit to warm up for 15 minutes before calibrating.
Quick Calibration
Restore:
Press all three keys
simultaneously until
[tc50] appears in the
display, press .
once and [`rSt] will
appear in the display.
Press and hold ß,
the display will show
[``no], press . to
change display to
[`YES]. Press and
hold . and ,
for 3 seconds to exit
the [`CAL] menu.
NOTE: Restore
Factory Calibration
[`rSt] = [`YES]
restores factory
calibration
values to all
calibration prompts.
Figure 8a -
Thermocouple Calibration
Figure 8b -
RTD Calibration
Calibration Menu
[tc50]
tc50
[tc00]
tc00
[``tc]
tc
[r`15]
r15
[r380]
r380
[`rSt]
rst
mV source = 50.000mV
Volt source = 5V
mV source = 0.000mV
Volt source = 0V
mV source = Temp. Compensation
Volt source = 0V
Decade Box = 15.00 ohms
Volt source = 0V
mV source = 380.00 ohms
Volt source = 5V
Store TC counts at 50.000mV
Store 5V remote set point
counts for use with TC
Store ambient counts
at 32° F. Type J.
Store low end RTD counts
Store 0V remote set point
counts for use with RTD
Store high end RTD counts
Store 5V remote set point
counts for use with RTD
Restore factory calibration
mV
source
Volt
source
S2
1
S1
+
2
3
+
45678
OT1IN OT2
POWER
Decade
Box
Volt
source
S2
1
S1
+
2
3
+
45678
OT1IN OT2
POWER
Store TC counts at 0.000mV
Store 0V remote set point
counts for use with TC