Owner's manual

[ Care and Use ManUal ]
BioSuite Peptide Analysis Columns 7
Note: Flow splitters after the column will introduce additional bandspread-
ing. System optimization, especially in a system that contains a flow split-
ter, can have dramatic effects on sensitivity and resolution. Optimization
includes using correct-depth ferrules and minimizing tubing diameter and
lengths. An example is given in Figure 9 where system optimization resulted
in a doubling of sensitivity and resolution of the metabolite in an LC/MS/MS
system.
Non-optimized LC/MS/MS System Optimized System
Figure 9: Chromatographic Performance on Non-Optimized vs. Optimized
LC/MS/MS System.
X. system modIFICatIon reCommendatIons For bIosuIte
2.1 mm I.d. peptIde analysIs Columns
1. UseamicroboredetectorflowcellwithBioSuite2.1mmi.d.Peptide
Analysis Column.
Note: Detector sensitivity is reduced with the shorter flowcell pathlength in
order to achieve lower bandspread volume.
2. Minimizeinjectorsampleloopvolume.
3. Use0.009inch(0.25mm)tubingbetweenpumpandinjector
4. Use0.009inch(0.25mm)tubingforrestofconnectionsinstan-
dardsystemsand0.005inch(0.12mm)tubingfornarrowbore
(>2.1mmi.d.)systems.
XI. addItIonal ConsIderatIons when usIng bIosuIte C
18
peptIde analysIs Columns
WatersBioSuitePeptideAnalysisColumnscontaineither3μm(i.e.,BioSuite
C
18
3μmPA-A)or3.5μm(i.e.,BioSuiteC
18
3.5μmPA-B)particlesthatpro-
vide very efficient separations without sacrificing column lifetime.
This section describes five parameters to consider when performing peptide
separations with BioSuite Peptide Analysis Columns.
Note: Columns that contain 3 μm or 3.5 μm particles have smaller outlet
frits to retain packing material. These columns should not be backflushed.
1. Backpressure: Backpressures for columns containing either 3 μm or
3.5 μm particles are higher than seen using columns packed with 5 μm
or 10 μm material.
2. Temperature:AsseeninTable1,useofacolumnheatersetat
elevatedtemperature(e.g.,40°C)willsignificantlyreducethe
observed system backpressure when using BioSuite Peptide Analysis
Columns.
3. Sampling Rate: Use a data sampling rate of 10 points per second or
greater.
4. DetectorTimeConstant:Useatimeconstantof0.1secondsforfast
analysis.
7.00 7.50
Non-optimized LC/MS/MS System Optimized System
8.00
7.00 7.50 8.00