Owner's manual
[ Care and Use ManUal ]
BioSuite Peptide Analysis Columns 3
2. ThedatainTable1wasgeneratedusingaBioSuiteC
18
, 3 μm PA-A
4.6mmx250mmanda2.1mmx250mmColumn.
• Proportionatelylesssystembackpressurewillbegeneratedusing
BioSuite C
18
, 3.5 µm PA-B columns that contain 3.5 μm material
compared to the 3 μm material contained in the BioSuite C
18
,
3 μm PA-A column.
• Proportionatelylesssystembackpressurewillbegeneratedusing
BioSuitePeptideAnalysisColumnsoflengthslessthan250mm.
3. RecommendedpHRange:2-8.
4. RecommendedOrganicSolvent/AdditiveConcentrationRange:
• BioSuiteC
18
3μmPA-A:0-100%acetonitrile(CH
3
CN)
Upto0.1%formicacid(FA)orupto0.025%trifluoroaceticacid
(TFA)
• BioSuiteC
18
3.5μmPA-B:0-100%CH
3
CN
Upto0.1%TFA
5. RecommendedTemperatureRange:20-50°C
Note: As indicated above, reduce flow rate when operating at ambient tem-
peratures may be required to avoid excessive column pressure on 250 mm
i.d. columns.
6. Recommended Cleaning Solvents/Procedures: A shift in retention or
resolution may indicate contamination of the column. Flushing with
aneatorganicsolvent(e.g.,100%CH
3
CNor80%Isopropanol/20%
CH
3
CN)isusuallysufficienttoremovethecontaminants.
7. Recommended Storage: For overnight storage, lower column tempera-
turetoambient(e.g.,from40°Cto25°C)andcontinuouslyflush
thecolumnwiththemobilephaseat10-20%ofthemaximumrecom-
mendedflowrate.Ifthecolumnisnottobeusedforseveraldays,
lowercolumntemperaturetoambientandstorethecolumnin100%
acetonitrile(CH
3
CN)withoutandeluentadditivessuchasformicacid
orTFA.Donotstorecolumnsinbufferedeluents.Ifyoumobilephase
contained a buffer salt, flush the column with 10 column volumes of
HPLCgradewatersbeforereplacingtheeluentwith100%CH
3
CN for
storage. Failure to perform this immediate step could result in precipi-
tationofthebuffersaltinthecolumnwhen100%CH
3
CN is introduced.
Completely seal the stored column to avoid evaporation and drying out
the bed.
GeneralConsiderationstoMaximizePerformanceofyourBioSuiteC
18
3 μm
PA-A and C
18
3.5 μm PA-B Columns.
II. sample preparatIon
1. Sample impurities often contribute to column contamination.
2. Itispreferabletopreparethepeptidesampleinthemobilephaseora
solventthatisweaker(lessorganicmodifier)thanthemobilephase.
3. Ifthepeptidesampleisnotdissolvedinthemobilephase,ensurethat
the sample, solvent and mobile phases are miscible in order to avoid
sample and/or buffer precipitation.
4. Filtersamplewith0.2μmmembranetoremoveparticulates.Ifthe
sample is dissolved in a solvent that contains an organic modifier
(e.g.,acetonitrile,methanol,etc.)ensurethatthemembranematerial
does not dissolve in the solvent. Contact the membrane manufacturer
withsolventcompatibilityquestions.
III. mobIle phase ConsIderatIons
Prevent air bubbles from entering the column during its installation, use,
and storage since this may cause degradation of column performance through
theformationofchannelsinthepackedbed.Mobilephasesmustbethor-
oughly degassed before use. This can be accomplished by vacuum filtration,
heliumspargingorbyin-linevacuumdegassing.Inadditiontodegassing
the solvent, vacuum filtration of the solvent will also prevent small particles
frompluggingthecolumnfrit.Youcanuse0.20µmor0.45µmmembrane
tofilteraqueousandaqueous/organicmobilephases.
Note: Consult with filtration membrane manufacturer for details on solvent
compatibility.
Note:Use high quality reagents, HPLC grade water, and HPLC grade solvents
for preparing eluents. The useful column lifetime is a function of numerous
factors including: the cleanliness and composition of the mobile phase and
the sample; the flow rate and pressure used; and the temperature. Refer to
the section below about “Cleaning” for information on extending column life.
Note: Cleaning, however, is not effective when the column is damaged by
irreversible sample adsorption, channeling, or exposure of the packing mate-
rial to excessive heat or shock.