User Manual

[ CARE AND USE MANUAL ]
ACQUITY UPLC CSH130 C
18
 Columns 1
Thank you for choosing a Waters ACQUITY UPLC
®
 CSH
™
 column. 
ACQUITY UPLC CSH130 C
18
, 1.7 µm and XP 2.5 µm columns feature 
Waters Charged Surface Hybrid (CSH) Technology which provides 
excellent peak shape, high efficiency and loading capacity for basic 
compounds when using acidic, low ionic strength mobile phases. 
This same particle technology is used in the XSelect
™
 CSH family of 
HPLC columns, thus enabling seamless transferability between HPLC 
and UPLC
®
 system platforms. The ACQUITY UPLC CSH130 packing 
materials were designed specifically for use with the ACQUITY UPLC 
systems and are manufactured in a cGMP, ISO 9001 certified 
manufacturing facility using ultra pure reagents. Each batch of 
ACQUITY UPLC CSH130 C
18
 material is tested chromatographically 
with acidic, basic, and neutral analytes as part of qualification for 
use in peptide mapping. The results are held to narrow specification 
ranges to assure excellent, reproducible performance. ACQUITY UPLC 
CSH130 C
18
 batches are also QC tested with a gradient separation 
of a tryptic digest of cytochrome c using 0.1% formic acid containing 
eluents. Finally, every shipped column is individually tested for packed 
bed efficiency and a Performance Chromatogram and Certificate of 
Batch Analysis are provided on the eCord
™
 intelligent chip. ACQUITY 
UPLC CSH130 C
18
 columns were designed and tested specifically for 
use on ACQUITY UPLC systems. ACQUITY UPLC CSH130 C
18
 columns 
will exhibit maximum chromatographic performance and benefits ONLY 
when used on holistically-designed ACQUITY UPLC systems since these 
systems and columns were created and designed to operate together. 
For these reasons, Waters cannot support the use of ACQUITY UPLC 
columns on any system other than an ACQUITY UPLC system.
CONTENTS
I. GETTING STARTED 
a. Column Connectors 
b. Column Installation 
c. Column Equilibration 
d. eCord Installation 
e. Initial Column Efficiency Determination 
f. VanGuard Pre-Columns
II. COLUMN USE 
a. Sample Preparation 
b. pH Range 
c. Solvents 
d. Pressure 
e. Temperature
III. COLUMN CLEANING, REGENERATING AND STORAGE 
a. Cleaning and Regeneration 
b. Storage
IV. INTRODUCING ECORD INTELLIGENT CHIP TECHNOLOGY
V. REPRESENTATIVE TEST CHROMATOGRAPH AND CONDITIONS 
 FOR SEPARATION OF PROTEIN DIGEST
VI. ADDITIONAL INFORMATION
ACQUITY UPLC CSH130 C
18
 1.7 µm AND XP 2.5 µm COLUMNS