Dosing and Mixing Unit Sidomix II USER MANUAL
Dear customer, Thank you for choosing a product from the Wassermann range. Wassermann DentalMaschinen incorporates the highest standards of quality and the latest technology. In order to enjoy maximum performance and years of trouble-free operation, please read this user manual carefully before you connect this device and start work, and operate the device according to the recommended guidelines.
Contents 1 Features............................................................................................................................4 2 Safety Symbols used in this Manual.................................................................................5 3 Safety Guidelines..............................................................................................................6 4 Responsibility for Operation or Damage...........................................................................
1 Features Mixing and dosing of addition-vulcanazing 1:1 two-component silicones. For even dosage of both components, both must have similar viscosity. The additional quantity flow control ensures an optimum mixing ratio. Should there be discrepancies, for example, for technical reasons (wear of the pump), or following a change of components with different viscosity, the mixing ratio can be adjusted.
2 Safety Symbols used in this Manual Warning! This is a warning of risk situations and dangers. Failure to observe this warning could be life-threatening. These warnings has to be observed. Information! This symbol draws your attention to specific features that has to be observed.
3 Safety Guidelines Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user manual. The equipment is designed only for its intended application. WARNING: Servicing and repairs should be carried out only by authorised specialists. Disconnect the power plug before starting any maintenance work. Make sure that the equipment is connected to the correct power source.
4 Responsibility for Operation or Damage The responsibility for operating the device lies exclusively with the owner or user if said device is incorrectly serviced, maintained or altered by persons not employed by an authorised dealer or if the device is used in a manner contrary to its specified purpose. The unit has to be maintained and operated in accordance with this user manual. Wassermann Dental-Maschinen GmbH is not responsible for damage arising from the nonobservance of these instructions.
6 Before Starting 6.1 Transport Before transporting the unit, ensure that it has been unplugged from the power socket. Make sure that it is packed correctly in order to avoid accidental damage. Be sure to check for any transport damage when unpacking the goods. Note down any damage if found. 6.2 Installation Open the box, remove the packing materials, and carefully lift out the device and accessories. Check the included accessories. The device has to stand horizontally on a steady and even surface.
7 Installation/ Start-up Before starting the unit, connect up the following: • Remove the closing plugs from the connections and dosage head. Finally, insert the power plug into the socket, making sure that the mains and the unit operate on the same voltage. 7.1 When using 2 Litre Material Containers (included): Fill the material container with the relevant component.
7.3 When using 6 kg Bulks: Remove the closings from the cans filled with two component silicone and screw the adaptor set onto the cans. Warning: Screw the colour marked adapter onto the can filled with coloured silicone compound – component B. Do not mix up! If you mix up the closings even for a short time, the device must be disassembled and cleaned. Insert the can with quick closings into the entry on the back of the device (the can will be upside down).
7.5 Replacing Bottles/ Cans: Remove the quick coupling/bottle adaptor from the device by pulling up the lock. Screw the quick coupling/bottle adaptor onto the new bottle/can and connect as explained above. Take special care to never confuse the bottles or connection hoses A+B – not even for a short time – since the silicone within the connections and pipes hardens instantly! The cleaning procedure is very demanding and may only be performed by authorised repair companies.
8 8.1 Operation General Operating Instructions All instructions for using the unit, whether in verbal or written form, are based on our own experience and experimentation and can only be regarded as guidelines. Switch on the device; the toggle switch lights green (right side of the unit). 8.2 Note for the Quantity Flow Control/ Setting of Quantity Flow Controller The adjusting screws are located on the side panel of the machine. (Behind the both-sided bores of the hood, please use a screwdriver.
8.3 Initial Start-up and Material Change * The unit will be delivered with a testing-liquid filled in! Observe at the first start-up * or at material change necessarily the following, because otherwise this will cause a mixing of the test liquid and the material or of "old" and "new" material: (Please pay attention to section 3 and Section 7.1 - 7.5! ) One-time procedure to clean up the hoses and the dosing pump: Place an empty bowl under the mixing head without mixing tip.
8.4 Using of the Mixing Tips and Dosage Mount the mixing cannula by turning it 90° on the mixing cannula adaptor (clean the adaptor before if necessary). Place the bowl or the model to be used for doubling the system under the mixing cannula. Press the red button: 1:1 mixed silicone runs out of the mixing cannula. We recommend collecting a small amount of the mixture initially output and just letting the substance output subsequently run into the form.
9 Troubleshooting Problem Cause Solution No silicone output. Hardened material within the old cannula. Mount a new mixing cannula. No silicone output. Mixing head clogged. Clean mixing head carefully. Warning: Silicone components must not get in contact with each other! No silicone output. A and B component was mixed up; silicone within the device system has hardened. The device needs to be disassembled and cleaned by a qualified technician. No silicone output.
10 10.1 Care and Maintenance Cleaning Disconnect the power plug before starting any maintenance work. The identification plate has always to be kept in easily legible condition and has not to be removed. Remove external dirt from time to time with some form of cold cleaner. Use only cold cleaners to avoid damaging the paintwork or removing the lettering. The equipment should be cleaned at regular intervals to ensure trouble-free operation.
10.4 Spare parts If necessary please contact our service hotline phone. 10.5 Service hotline 0049 (0)40 / 730 926 -20/ -24 10.6 Scope of Delivery/ Accessories Included parts Set material cask incl. quick fitting coupling, 2 x 2 liter Mixing canulas, pink, 10 pcs. Accessories Set material cask with quick fitting coupling, 2 x 2 liter Stand small for duplicating silicones, 2 x 6 kg-cask Stand tall for duplicating silicones, 2 x 25 kg-cask Connecting tubes cpl., 2 pcs.
12 Disposing of the Unit The unit has to be disposed by an authorized recycling operation. The selected company has to be informed of all possibly health-hazardous residues in the unit. 12.
13 EU Declaration of Conformity in accordance with 2014/35/EU (Low Voltage Directive) and 2014/30/EU (Electromagnetic Compatibility Directive) and 2006/42/EC (Machinery Directive) and 2011/65/EU (RoHS Directive) Manufacturer: WASSERMANN Dental-Maschinen GmbH Rudorffweg 15-17 21031 Hamburg Germany Model: Sidomix II Sidomix II 115 V Applicable standards: DIN EN 61010-1 DIN EN 61000-6-3 DIN EN 61000-6-1 DIN 45635-1 DIN EN 60335-1 Product description: Mixing and Dosing unit, 1:1, for dental application