Specifications

42
Display
Units
Program Controler
Tolerance judgment result output
Serial BCD output
Analog output
Serial BCD output
Analog output
Digimatic output
Personal computer
Digimatic Mini-processor
DP1-VR (264-504)
LGB
LGF
LGK
LGF with
Origin Point Mark
EB-11D
EB-11Z
EB-11P
LGD
LGS
External switch box
FEATURES
SPECIFICATIONS
EB Counter
DIN size (96 x 48mm) assembly-type display unit
FEATURES
• Possible to produce 3-step/5-step x 7 kinds of tolerance output and limit value output independently for each of 7 channels.
• Provided with serial BCD output capability, which makes the connection to a programmable controller or personal computer, etc., possible with
the minimum cabling requirement.
Possible to perform dynamic measurement with the simplified analog output.
542-092-2 542-094-2 542-093-2
For differential square-wave output
gage head
For differential square-wave output
gage head with Origin Point Mark
For Digimatic code output gage head
Order No. 542-092-2 542-094-2 542-093-2
Model No. EB-11P EB-11Z EB-11D
Number of gage inputs 1
Maximum input frequency 1.25MHz*
1
(differential square-wave) Depends on the gage head connected
Resolution 0.01mm, 0.005mm, 0.001mm, 0.0005mm, 0.0001mm 0.01mm, 0.001mm
Tolerance judgment display LED display (3 steps: amber, green, red / 5 steps: amber, amber flashing, green, red flashing, red)
Interface
Serial BCD Bit-serial format, open-collector
Analog output 2.5V + Counting value X voltage resolution (25mV/2.5mV): Full-scale 0 to 5V
Digimatic input/
output
• Connecting to the external switch box (02ADF180) makes it easy to enter tolerance limits and preset values.
Note) This cannot be used when the gage is connected to a Mitutoyo DP-1VR Digimatic Mini-Processor
• Possible to connect to a Mitutoyo DP-1VR Digimatic Mini-Processor.
• Number of tolerance steps can be expanded by using multiple EB-D counters.
Output
Tolerance judgment L1 to L5, open-collector
Control signal Normal operation signal (NORMAL), open-collector
Input Control signal
Preset, display hold, peak value clear, tolerance judgment BANK switch, open-collector or no-voltage contact signal (with/without contact
point)
Power supply voltage DC+12 to 24V
Power consumption 6W (500mA) or less (Secure power supply more than 1A for each unit.)
Operating temperature 0ºC to 40ºC (20%RH to 80%RH, without condensation)
Dimensions W96 x H48 x D156mm
Optional accessories
02ADB440:Output connector with cover, 02ADN460:AC adapter, AC adaptor connecting cable,Connecting cable,
02ADF180: External switch box
Compatible gage heads LGK, LGB, LGF LGF with Origin Point Mark LGD, LGS, ID-C
Mass 400g 400g 400g
*1: Depends on the gage head connected.