Specifications
Appendix B. Configuration Table MI 611-211 – October 2005
53
5CONFIG BUFFERS- - - If pH
BUFFERS AMERICAN,
NIST,
EUROPEAN,
SPECIAL
AMERICAN
SPECIAL BUFFER 1 - - - If SPECIAL
NUM PTS 02 AND 11 02
TEMPx -99.9 and 999.9 0.00
°C or °F
VALx -2.00 and 14.00
pH
0.00
SPECIAL BUFFER 2 - - -
NUM PTS 02 AND 11 02
TEMPx -99.9 and 999.9 0.00
°C or °F
VALx -2.00 and 14.00
pH
0.00
SPECIAL BUFFER 3 - - -
NUM PTS 02 AND 11 02
TEMPx -99.9 and 999.9 0.00
°C or °F
VALx -2.00 and 14.00
pH
0.00
CONFIG PASSCODE - - -
PASSCODE MINOR - - -
MINOR 0 AND 9999 0800
PASSCODE MAJOR - - -
MAJOR 0 AND 9999 0800
CONFIG STRINGS - - -
STRINGS TAG NUM - - -
TAG NUM
STRINGS TAG
NAME
- - -
TAG NAME
STRINGS LOCATE - - -
LOCATE
STRINGS DEV
NAME
- - -
DEV NAME
CONFIG AC PWR - - -
6
AC PWR
60, 50 Hz 60
CONFIG DIAGS - - -
Structure
Diagram Prompt/Parameter
Parameter
Limits
Factory
Configuration
User
Configuration
Remarks