Specifications
MI 611-211 – October 2005 3. Operation
28
The message may be enabled, disabled, or cleared. The following Foxboro sensors contain a
preamplifier: 871A-4, 871PH-3, 871PH-4, PH10-.P and ORP10-.P.
ATCSHORT, ATC OPEN
This message reports a problem if the resistance of the temperature compensator is greater or less
than the expected resistance of the device configured. The message for each may be enabled,
disabled, or cleared.
COAT
This message reports an increase in reference junction resistance. This can result from a fouled
reference junction or if the electrode is not immersed in the process solution. The diagnostic
checks the resistance against solution ground. An error message appears if the resistance is greater
than the user set limit. The limit may be set between 0 and 100 kΩ in increments of 10 kΩ. The
default value is 80. The message may be enabled, disabled, or cleared.
NOTE
1. During normal operation, this diagnostic may indicate that the pipe or tank has
gone dry and the sensor is no longer in solution.
2. During sensor maintenance such as cleaning or calibration, this diagnostic may
appear during the times the sensor is out of solution. When the sensor is again
immersed, the diagnostic should turn itself off in a few minutes. If the diagnostic
remains on, then the sensor should be treated as having a coating.
AGING
This message reports an aging glass electrode. An error message appears after a 2-point calibration
in pH buffer solutions when the Nernst slope of the sensor has decreased sequentially five times.
The message may be enabled, disabled, or cleared.
4-20 RNG
This message reports that the measurement that is tied to the analog output is outside the range
that is configured for the application. The message may be enabled, disabled, or cleared.
COMP RNG
This message reports that the measured temperature or the absolute measurement is outside the
temperature or chemical compensation curve that is configured for the application. The message
may be enabled, disabled, or cleared.
MEAS RNG
This message reports that the measurement cannot be displayed accurately in the precision/scale
that is configured for the application. The message may be enabled, disabled, or cleared.
Table 11 shows what diagnostics are available with various Foxboro sensors if enabled and
configured with reasonable limits.