Specifications

MI 611-211 – October 2005 3. Operation
16
NOTE
Corrective action for AGING error is to check the value of the slope in the Status
menu. The glass or ISE electrode may be reaching the end of its useful life if the
SLOPE differs greatly from the theoretical value (80% is usual limit but application
requirements should dictate lowest acceptable slope).
Status
(Refer to “Structure Diagram 8” on page 45.)
The Status mode enables you to view the measurement and system parameters and thus assess the
performance of the loop. You may not edit them in this mode. For a string display of more than
eight characters, use the Left and Right arrow keys to scroll the display. All measurements
(outputs) continue while in the Status mode. To enter the Status mode, press the STATUS key.
Move through the menu by using the Up and Down arrow keys. While you are in the Status
mode, two segments of the bar graph will appear under the legend STATUS at the top of your
front panel. The Status menu contains the following parameters.
Configuration Errors
CAL REQ Calibration required Calibrate
4 - 20 RNG Measurement outside configured range Reconfigure range or correct
process error
MEAS RNG Measurement outside configured precision or scale Reconfigure precision or
correct process error
COMP RNG Measurement outside compensation curve Reconfigure compensation
or correct process error
Table 5. Status Parameters
Parameter Description
TEMP Process Temperature
SLOPE Efficiency of sensor expressed as the deviation in percent of the slope of the most
recent calibration compared to the standard (e.g., mV/pH/59.16 mV/pH)
ABSOLUTE Uncompensated millivolt signal from sensor
(a)
mA Analog Output: Analog output value in mA
Digital Output: Output shown on secondary display in mA.
GLASSRES Indicates whether resistance of glass electrode is acceptable
ASYMM Asymmetry potential (mV difference from the expected theoretical isopotential pH
voltage of zero)
REF RES Resistance of reference electrode
ATC RES Resistance of temperature compensator
CAL NAME Name of last calibrator
LAST CAL Date of last calibration (mm/dd/yy)
Table 4. Diagnostic Error Messages (Continued)
Message Description Corrective Action