Specifications
MI 611-211 – October 2005 1. Introduction
4
Specifications
Sensors: Virtually any glass or antimony pH, ISE, or noble metal ORP sensor including the
following:
PH10 DolpHin™ Series pH Sensor
ORP10 DolpHin™ Series ORP Sensor
871A pH/ORP Sensor (see Note)
871PH pH/ORP/ISE Sensor (see Note)
EP459A Fluoride Sensor
EP460 pH/ORP Sensor
EP462 pH Sensor
EP466 pH Sensor
NOTE
871A-2, 871PH-1, and 871PH-2 Sensors, although compatible with 870ITPH
Transmitters, do not supply full sensor diagnostic capability.
Maximum Distance Between Sensor and Transmitter:
With preamplified sensor: 152 m (500 ft)
With non-preamplified sensor that utilizes a driven shield
(PH10-.N, 871A-1, -3; 871PH-5, -6): 15.2 m (50 ft)
Temperature Compensation Inputs:
100 ohm platinum RTD, 2- and 3-wire
1000 ohm platinum RTD, 2- and 3-wire
Balco 3000 ohm RTD
Ambient Temperature: -25 to +55° C
Relative Humidity: 0 to 95% noncondensing
Electromagnetic Compatibility (EMC): The Model 870ITPH Intelligent Transmitter complies
with the requirements of the European EMC Directive 89/336/EEC when the sensor cable and
I/O cables are enclosed in rigid metal conduit. (See Table 2.)
Power Supply Requirements (for 4 to 20 mA output): 12.5 to 42 V dc (see Figure 2)
Load Limitations (for 4 to 20 mA output): 0 to 1450 ohms (see Figure 2)