Instruction manual

127
Output Mode
Click on the radio buttons to manually control the relay. The normal selection will be Auto, where
the controller uses the various set points to open and close the relay. You can force the relay closed
regardless of the set points by selecting Hand, or force the relay open regardless of the set points
by selecting Off.
If you select Off, the relay will not activate again until either Hand or Auto is selected (or if power
is cycled)! If you select Hand, the relay will activate until either Auto or Off is selected, or until
the Hand Time Limit (programmed in the next menu) expires or until the controller power is
cycled.
The Off mode is handy to prevent the relay from activating while you are changing set points. The
Hand mode is handy to be able to test the pump or valve, wiring, etc without needing to change set
points.
Click on the “Submit” button to save the changes.
Event Log
Click on the link to view the event log. The event log contains the date and time of each relay
activation and deactivation, as well as the state of each Interlock, Level Switch, or Generic digital
input.
Type of Event
Status to generate a ‘1’ in the log
Status to generate a ‘0’ in the log
Relay output
Relay activated
Relay not activated
Interlock DI
In the Interlock state as defined in
the Interlock page
In the non-Interlock state as defined in
the Interlock page
Level Switch DI
In the Low Alarm state as defined
in the Level Switches page
In the Normal state as defined in the
Level Switches page
Generic DI
Alarm selected
In the Alarm state as defined in the
Generic Inputs page
In the Normal state as defined in the
Generic Inputs page
Generic DI No
Alarm selected
Switch closed
Switch open
A window will open to ask if you would like to open the file or save it to disk. Select the method
you would like to use. The file is .csv (comma separated variable) file that will open in a
spreadsheet such as Excel.
6.37 Analog Output (1-4) Menu Retransmit Mode
(only appears if one or more 4-20 mA output option cards are installed)
4-20 mA Output Control Mode
Click on the pull-down to select between Not Used , Proportional Feed and Retransmit control
mode. This section assumes that the control mode is set to retransmit.
Input Assignment
Click on the arrow to pull down the choices of which sensor measurement will be retransmitted as
a 4-20 mA signal. The choices are any of the sensor inputs, any of the sensor’s temperature
compensation inputs, LSI, RSI, any paddlewheel flow meter or any 4-20 mA input signal (only if
the 4-20 mA input option is installed).