Instruction manual

33
The “1pt Buffer Cal” is not as accurate, since it assumes that the slope of the electrode is 59
mV/pH unit, which will only be true with a new electrode. This can be OK if the pH never varies
far from the value of the standard solution used in the calibration. This method uses manual
temperature compensation, since the temperature sensor is contained in the system conductivity
sensor, which is not placed in the buffer solution.
The third is a “1 pt Process Cal”, where the pH electrode is left in the process solution, and the pH
value is entered to match that of a hand held pH meter, or litmus paper. It also assumes a perfect
slope. Automatic temperature compensation is used, since the pH electrode is installed in the
system water with the conductivity sensor.
2 pt Buffer Cal
Step 1 will explain that control is suspended during the calibration procedure. Press the Enter key
when the cursor is on “Continue” to start the calibration. To cancel the calibration, press the Next
key to move the cursor to “Cancel”, then press the Enter key.
Step 2 will ask you to enter the temperature of the buffer solutions. Use the Arrow keys to change
the value of the temperature, then press the Enter key to highlight ‘CONTINUE’, then press Enter
again to put the value into memory.
Step 3 will prompt you to rinse the electrode, and place it into the first buffer solution. Press the
Enter key to continue.
Step 4 will ask you to enter the value of the first buffer solution. Use the Arrow keys to change the
value, then press the Enter key to highlight ‘CONTINUE’, then press Enter again to put the value
into memory.
Step 5 will display the mV output of the electrode in the first buffer. When this reading is stable,
press the Enter key to continue.
Step 6 will prompt you to rinse the electrode and place it in the second buffer. Press the Enter key
to continue.
Step 7 will ask you to enter the value of the second buffer. Use the Arrow keys to change the
value, then press the Enter key to highlight ‘CONTINUE’, then press Enter again to put the value
into memory.
Step 8 will display the mV output of the pH electrode in the second buffer. When the reading is
stable, press the Enter key to continue.
At the end of the calibration, the display will tell you if the calibration was successful or if the
calibration failed. If successful, press the Enter key after the electrode has been replaced into the
sample stream to resume control. If failed, refer to Section 7 for troubleshooting help. The
calculated slope being more than 80% different from theoretical causes a calibration failure, or by
the pH 7 mV offset being more than 60 mV.
1 pt Buffer Cal
Step 1 will explain that control is suspended during the calibration procedure. Press the Enter key
when the cursor is on “Continue” to start the calibration. To cancel the calibration, press the Next
key to move the cursor to “Cancel”, then press the Enter key.
Step 2 will ask you to enter the temperature of the buffer solution. Use the Arrow keys to change
the value of the temperature, then press the Enter key to continue.