Instruction manual

116
Faulty conductivity sensor wiring. Make sure that the sensor cable is connected (if
system conductivity) properly, and there are
no shorts in any extension cables. Also make
sure that the sensor cable is not routed near any
AC voltage wiring. Make sure that the cable is
properly shielded.
Faulty conductivity sensor. Replace sensor.
pH Sensor Error
This error message is caused by the signal from the pH electrode being outside of the normal range
(pH less than –2 or greater than 16, or input voltage outside 1.45 VDC to +1.45 VDC). The pH
output relay will be forced off.
Possible Causes Corrective Actions
Faulty pH cartridge. Replace cartridge.
Faulty pH electrode wiring. Make sure that the electrode cartridge is firmly
screwed into the housing. Make sure that there
is no water inside the housing. Make sure that
the +/- 5 VDC and the IN-, IN+ wires are
firmly attached.
Faulty preamplifier housing. If the controller passes self test, the wiring is
correct, the +/- 5 VDC signal is present and
the cartridge is attached and OK, replace the
housing.
Faulty sensor option circuit board. Go to the pH Input Menu and perform a self
test. If it fails, remove the electrode wires f
rom the pH input option board and try the self
test again. If it passes, the problem is in the
electrode, housing or its wiring. If it still fails,
replace the pH input option board.
ORP Sensor Error
This error message is caused by the signal from the ORP electrode being outside of the normal
range (-1450 to 1450 mV). The ORP output relay will be forced off.
Possible Causes Corrective Actions
Faulty ORP cartridge. Replace cartridge.
Faulty ORP electrode wiring. Make sure that the electrode cartridge is firmly
screwed into the housing. Make sure that there
is no water inside the housing. Make sure that
the +/- 5 VDC and the IN-, IN+ wires are
firmly attached.
Faulty preamplifier housing. If the controller passes self test, the wiring is
correct, the +/- 5 VDC signal is present and
the cartridge is attached and OK, replace the
housing.