Technical data

33
W&T
Subject to errors and modifications
3.2 Calibration
The sensor can be calibrated using single-point and two-point
reference measurements and corresponding entries for offset
values.
In single-point compensation the entered offset value is added
to the measured value, whereas in two-point compensation a
straight line is calculated for compensating the entire
measuring range. The offset is allowed to be max. 20% of the
total range. The offset may be negative, but the full scale values
may not fall below the 0% mark.
For offset you enter the value to be added to the measured
value in order to reach the desired value.
To retain calibration setting, the user can store a comment text.